Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | Patient Monitor |
Still deciding? Get samples of US$ 500/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Features:
1. 12.1" color TFT display, high resolution: 800x600dpi;
2. Separated para module: ECG, NIBP, SpO2 module;
3. Real time S-T segment analysis, and ARR analysis;
4. SpO2 Pitch Tone Variation and Drug Dose Calculation;
5. Parameter waveform and characters color selectable
6. Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic 7-lead ECG waveforms
7. Standby mode for Energy saving and suspend monitoring;
8. Capture dynamic waveforms;
9. Internal storage: 3600s full-disclosure waveform review;
10. External storage: SD card slot enables memory extension;
11. Built-in rechargeable lithium battery, and battery volume checking,
12. VGA Output; AC power checking
13. Efficient resistance to interference of defibrillator and HF knife;
14. High accuracy SpO2 &NIBP, workable for neonate;
15. Networking capacity and TCP/IP networking platform enable abundant future upgrade
16. Time delay switch-off function keep the monitor data restart monitor;
17. Complete and flexible mounting solution for transportation and in-patient monitoring
Standard Parameters: 5-lead ECG, SpO2(F-5),NIBP, TEMP,RESP,PR,HR.
Optional Parameter: Touch Screen, Thermal Recorder, 3/12-lead ECG, 2/4-IBP,4-TEMP,2-Sp02,Masimo SpO2,BIS, Phaisein/ Respironics EtCO2, ICG/CO, FHR, FM, TOCO,
Standard parameters:
Ecg:
Lead selection: five lead standard; Up to 7 leads can be displayed in the same screen I, II, III, AVR, AVL, AVF, V chest lead ECG detection range: 0-350BPM
Resolution: 1bpm Accuracy: ±1%
Ecg amplification width: 0.05HZ-100 Hz S-T segment monitoring: -2.0mV to 2.0mV arrhythmia analysis: yes
Electrode off indication: sound, light prompt
Scanning speed: 6.25, 12.5, 25, 50mm/ SEC Gain selection: ×1/2, ×1, ×2, ×4
Ecg calibration: 1mV+5%
Protection and isolation withstand 4000V/AC/50Hz voltage; Ecg cable guide connection:
Standard configuration: adult electrocardiogram cable
Optional configuration: special ECG cable for children and newborns
Alarm: alarm value, lower limit can be set, automatic memory; There is alarm review;
Blood oxygen saturation:
Display: Blood oxygen number, pulse chart, waveform, pulse value Blood oxygen range: 0%-100% Adult/child/newborn resolution: 1%
Accuracy: ±2% (not defined below 70%) Pulse rate:
Measurement range: 0 ~ 2501bpm Pulse accuracy: ±1bpm
Alarm range: upper limit 80 ~ 2501bpm
The lower limit is 20 ~ 1501BPM
Sensors:
Standard configuration: Adult soft ring
Optional configuration: children's soft hand ring, newborn bundle alarm value: upper and lower limits can be set, automatic memory
Non-invasive blood pressure:
Measurement method: intelligent oscillation method
Measurement parameters: systolic blood pressure, diastolic blood pressure, mean blood pressure, pulse rate Working mode: manual, automatic, continuous measurement
Unit: mmHg/kPa Optional
Automatic measurement interval: 0 to 250min adjustable overvoltage protection: software and hardware overvoltage protection
Measurement range: adult: 10mmHg ~ 300mmHg Children: 10mmHg ~ 200mmHg neonates 10mmHg ~ 155mmHg
Cuff pressure range: 0-320 mmHg
Cuff:
Standard configuration: Adult cuff
Select configuration: newborn cuff, child cuff cuff inflation time: less than 6 seconds
Test venting time: less than 15 seconds Alarm range:
Upper systolic blood pressure is 120-240 MMHG
The lower limit of 40-150 mmHg
The upper limit of average pressure is 70-200 mmHg
The lower limit of 20-120 mmHg
The upper limit of diastolic blood pressure is 50-80 mmHg
The lower limit of 10-80 mmHg
Temperature:
Temperature range: 0 ° C to 50 ° C Temperature measurement error: ±0.1 ° C
Resolution: 0.1ºC Accuracy: ±0.1 ºC Probe:
Standard configuration: permanent body temperature probe
Optional configuration: esophageal temperature probe, rectal temperature probe. Alarm value: upper and lower limits can be set, automatic memory
Alarm: IT CAN SET alarm for all monitoring parameters, and has alarm device that sends out sound and light, and can cancel alarm
Breathing:
Measurement method: chest impedance method
Respiratory rate (RR) : A. Measurement range: 0-120bpm
B. Measurement accuracy: ± 1BPM or 5%bp
Alarm: Alarm value can be set, automatic memory